== Sample preparation  ==
== Sample preparation  ==
-
Material:&nbsp; Palmoplantar stratum corneum scrapings from healthy volunteers
+
Material:&nbsp; Palmoplantar stratum corneum scrapings from healthy volunteers  
-
Extraction:&nbsp; Chloroform–Methanol (1:1, v/v)
+
Extraction:&nbsp; Chloroform–Methanol (1:1, v/v)  
-
 
+
-
== Instrumentation and  method ==
+
 +
== Instrumentation and method  ==
=== Pump  ===
=== Pump  ===
-
*Type: Spectra System P 4000&nbsp;&nbsp;(&nbsp;Thermo separations)
+
*Type: Spectra System P 4000&nbsp;&nbsp;(&nbsp;Thermo separations)  
*Mode: Gradient  
*Mode: Gradient  
*Solvent(s): A: chloroform; B: chloroform, n- propanol,&nbsp;acetic acid (80:20:2)&nbsp;  
*Solvent(s): A: chloroform; B: chloroform, n- propanol,&nbsp;acetic acid (80:20:2)&nbsp;  
Line 45: Line 46:
=== Mass spectrometer  ===
=== Mass spectrometer  ===
-
*Type: Finnigan LCQ classic (ion trap mass spectrometer)
+
*Type: Finnigan LCQ classic (ion trap mass spectrometer)  
-
*Ionization mode: APCI, positive full scan mode
+
*Ionization mode: APCI, positive full scan mode  
*Source temperature(s):&nbsp; APCI heater:&nbsp; set to 500°C,&nbsp; heated capillary:&nbsp;150°C<br>
*Source temperature(s):&nbsp; APCI heater:&nbsp; set to 500°C,&nbsp; heated capillary:&nbsp;150°C<br>
 +
<br>
 +
<br>
-
 
+
<br>
-
 
+
-
 
+
== Reference  ==
== Reference  ==

Revision as of 12:17, 2 September 2008

How to use:

  • Click 'Edit' and copy the source code to clipboard
  • Create new page/article and paste the source code into your new article.
  • Modify as needed.


Contents

Sample preparation

Material:  Palmoplantar stratum corneum scrapings from healthy volunteers

Extraction:  Chloroform–Methanol (1:1, v/v)

Instrumentation and method

Pump

  • Type: Spectra System P 4000  ( Thermo separations)
  • Mode: Gradient
  • Solvent(s): A: chloroform; B: chloroform, n- propanol, acetic acid (80:20:2) 
  • Gradient:
Time [min] Flow [ml/min]  % Solvent A  % Solvent B
0 1 100 0
15 1 0 100

Column

  • Type: silica column LiChroCart 125-4 with Si 60 (5µm) (merck)

Mass spectrometer

  • Type: Finnigan LCQ classic (ion trap mass spectrometer)
  • Ionization mode: APCI, positive full scan mode
  • Source temperature(s):  APCI heater:  set to 500°C,  heated capillary: 150°C




Reference

Hany Farwanah , Peter Nuhnb, Reinhard Neubert , Klaus Raith. Normal-phase liquid chromatographic separation of stratum corneum ceramides with detection by evaporative light scattering and atmospheric pressure chemical ionization mass spectrometry. Analytica Chimica Acta 492 (2003) 233-239.